Which scales may be used in a medical facility

visibility 2476 Views person Posted By: SKLEP DLA LEKARZA list In: Charder medical scales
Which scales may be used in a medical facility

Read an interpretation of the legal provisions on this question, prepared by a law firm.

In response to the question whether scales used in medical facilities are subject to the Polish Measures Act, we explain as follows.

First, scales used in medical facilities (providing healthcare services) must be medical devices within the meaning of the Polish Medical Devices Act of 7 April 2020 (Journal of Laws of 2022, item 974), because, a contrario, these scales are excluded from the Act on conformity assessment and market surveillance systems (Article 1(5) of that Act).

Next, under Article 8(1)(1) of the Polish Measures Act of 11 May 2011 (consolidated text: Journal of Laws of 2022, item 2063, as amended), measuring instruments that may be used to protect health, life and the environment and are specified in regulations issued under paragraph 6 are subject to legal metrological control.

Sharing the view of Zdzisław Wąsowicz, Marek Antosiuk and Marek E. Dębowski (Regional Office of Measures in Białystok) (https://www.gum.gov.pl/ftp/pdf/Biuletyn/Artykuly/3-4_2016/53-58_Prawna_kontrola_metrologiczna_wag_nieautomatycznych_elektronicznych.pdf) *, it should be stated that electronic non-automatic weighing instruments in EU member states are subject to Directive 2014/31/EU of the European Parliament and of the Council of 26 February 2014 on the harmonisation of the laws of the Member States relating to the making available on the market of non-automatic weighing instruments (NAWI) (OJ L 96, 29.3.2014, p. 107). This Directive (module B) covers non-automatic weighing instruments:

  • with digital indication,
  • platform scales,
  • tank scales,
  • electronic vehicle scales,
  • hybrid vehicle scales.

Therefore, every non-automatic weighing instrument placed on the market or put into use after 1 May 2004 and used to determine mass:

  1. in trade,
  2. as a basis for calculating market fees, customs duties, taxes, bonuses, discounts, penalties, remuneration, compensation or similar charges,
  3. when applying legal provisions and when experts give opinions in court proceedings,
  4. of patients in medical practice for monitoring, diagnosis and treatment,
  5. when preparing prescription medicines in pharmacies,
  6. in analyses carried out by medical and pharmaceutical laboratories,
  7. when packaging goods, must undergo a conformity assessment procedure.

Such instruments carry the CE mark and the letter "M" with the year and the notified body number, confirming their conformity with the NAWI directive. Using electronic non-automatic weighing instruments in the areas listed in points 1-7 that have not undergone conformity assessment and do not carry these markings, required for the verification of scales, exposes users to fines imposed by state authorities.
Scales for any other purpose are not subject to NAWI conformity assessment. However, their rating plates must show the manufacturer's name, maximum capacity, scale interval and the CE mark confirming that the product complies with general safety rules (meets the EMC and LVD directives).

In view of the above, there is no doubt that scales used by healthcare providers to determine a patient's body weight must be verified under the NAWI Directive and must be medical devices within the meaning of the Polish Medical Devices Act of 7 April 2020 (Journal of Laws of 2022, item 974).

--
* The validity periods of verification of non-automatic weighing instruments have changed (all physician scales, including electronic ones, are non-automatic weighing instruments): for accuracy classes III and IIII, among others, it is now 25 months.